Dosimetric study of simultaneous integrated boost intensity-modulated radiotherapy for cervical and upper thoracic part esophageal carcinoma

蒋俊,权循凤,洪浩,李锐
DOI: https://doi.org/10.19405/j.cnki.issn1000-1492.2014.03.029
2014-01-01
Abstract:Seven cervical part and upper thoracic part esophageal carcinoma patients were respectively designed for 3 ,5 ,7 fields in simultaneous integrated boost intensity-modulated radiotherapy plans to compare the differences in target and organ at risk dose. Results revealed that the target area dose coverage,conformity index and homogeneity index in 3 fields plans were inferior to 5 fields and 7 fields plans. Moreover, the both lung V5 in 7 fields plan was greater than 5 fields and 3 fields plans. The target area could receive sufficient dose in 3 fields plan, but the uni-formity and conformity were unsatisfied. 5 fields plan made the target area receive ideal dose distribution, which showed 5 fields plan was the first choice in clinic.
What problem does this paper attempt to address?