In Vitro Accuracy Study of a Implanting NavigationTemplate

Min Peng,Wei Fei
DOI: https://doi.org/10.16252/j.cnki.issn1004-0501-2016.11.005
2016-01-01
Abstract:Objective Personalized implant surgical guide based on CBCT is also called CBCT navigation implant tem-plate,having important significance upon accurate placement of implants. The objective of this study is to make a simple CBCT guide and evaluate its accuracy. Methods A mandibular model was reproduced and a plastic guide plate was hand-tailored. The guide plate was embedded with 4 metal beads as radiation markers and placed at the occlusal surface of the mandibular model. The model with surgical guide was photographed using CBCT. Simulated implanting was processed on the CBCT images using virtual im-plants,so three-dimensional space position data of virtual implants with respect to guide plate could be collected. The implanting guide holes were manufactured on the surgical guide plate according to the position data using specifically designed 5 axis drilling ma-chine. The well-prepared surgical guide was placed on the lower jaw and photographed by CBCT again. The axial difference between axle wires of implanting guide holes in second imaging and axes of virtual implants in first imaging was evaluated. Results The error of enter-point and the difference of axial angle were(0. 47 ± 0. 27)mm,(0. 57 ± 0. 23)mm and(0. 64 ± 0. 16)°,(0. 57 ± 0. 15)°,re-spectively. Conclusion In this study,the accuracy of CBCT guided implanting template was high and the error of 20 degree en-ter-point was larger than 10 degree enter-point at each locus.