Clinical Application And Evaluation of Galileo Echo Automatic Blood Analyzer

Jing LIANG,Wen LIU
DOI: https://doi.org/10.14033/j.cnki.cfmr.2016.1.028
2016-01-01
Abstract:Objective:To evaluate the application of Galileo Echo automatic blood analyzer on blood grouping and irregular antibody screening tests. Method:ABO,Rh(D) blood grouping and the irregular antibodies were studied in 9082 samples using Galileo Echo automatic blood analyzer,at the same time, comparing with manual method(test tube method).Result:The accuracy of Galileo Echo automatic blood analyzer in ABO and Rh(D) blood grouping identification were 99.75%(9059/9082) and 99.88%(9071/9082),there was no significant difference in the manual method.The positive rate of the irregular antibodies was 0.54%(49/9082).Conclusion:The results of Galileo Echo automatic blood analyzer in ABO,Rh(D) blood grouping and the irregular antibodies tests are reliable, its experimental operation is automatically,quickly and convenient.Moreover,the experimental results can be permanently preserved,which is suitable for clinical blood grouping tests.