Research on Auto Vue Innova Automatic Instrument for Pre-transfusion Testing

王敏,王保龙,姚萍,蒋光明
DOI: https://doi.org/10.3969/j.issn.1671-2587.2011.02.009
2011-01-01
Abstract:Objective To explore Auto Vue Innova automatic instrument for identification of blood groups and unexpected antibody screening.Methods Forward and reverse ABO/Rh typing and unexpected antibody screening were performed on 9 652 samples using the Auto Vue Innova automatic instrument and manual method,then a comparison was made.Positive samples would be confirmed by conventional indirect anti-globulin technique.Results The rates of first accurate ABO typing by automatic instrument and tube method were 99.71%(9 624/9 652) and 99.75%(9 628/9 652) respectively.There was no significant difference between the two methods.The rates of first accurate RhD typing by the two methods were all 100.00%.The rates of unexpected antibody detection and the number of false positive were 0.47%(45/9 652),7 and 0.41%(40/9 652),2 respectively.The throughput per hour using Auto Vue Innova automatic instrument and manual method were 96,145 and 21,48 respectively.Conclusion The automatic instrument is accurate,safe and has higher sensitivity.It is easy to inquiry and quote and is suited for batch operations.