The Research of Mathematical Models in BlPM/ClPM Key ComParison CCM. FF-K4. 2. 2011 Volume ComParison at 100 μL

Lin TONG,Jin-tao WANG,Zi-yong LIU,Kai WEI,Long ZHANG,Li-gong GUO,Xiao-hui ZHAO
DOI: https://doi.org/10.3969/j.issn.1000-1158.2015.02.10
2015-01-01
Abstract:According the key comparison found that the atmospheric pressure has a great impact on the pipette calibration results,and after pressure correction there is a better consistency in report. On the other hand the micropipette reliability is the major source of uncertainty on the total calibration uncertainty through a lot of laboratory tests. Finally the mathematical models are proven to be accurate and reliable through a numerical example based on the key comparison,and they could be used in calibration and data processing for micropipettes.
What problem does this paper attempt to address?