Thermodynamical analysis of the effect of environmental conditions on measurement error of bell prover

Yiping LIU,Chao CHEN,Jinchuan WU,Xiaojun QUAN
DOI: https://doi.org/10.11857/j.issn.1674-5124.2017.06.025
2017-01-01
Abstract:Using air conditioning system to adjust the environmental temperature of bell type gas flow standard device, but its control effect is difficult to fully meet the requirements of the verification of high accuracy flowmeter. In order to obtain time-variant state parameters during the operation period of bell prover, the discharging process was regarded as a one-exit variable-volume system which was similar to clearance -sealed piston pneumatic system, and a thermodynamic model was presented for the displacement and the velocity of bell prover, as well as the pressure and the temperature inside the bell. An analysis of the thermodynamic model was implemented by solving four first order differential equations. The results show that neglecting the heat exchange between gas and environment will give rise to measurement error, which will become larger with the increased temperature difference. The random variation of environmental pressure and temperature caused by air conditioning system prevents the inner pressure from reaching constant, thus the measurement error caused by variable flow rate is hardly eliminated by the improvement of mechanical structures.
What problem does this paper attempt to address?