Study of Field Uniformity Measurement and Dose Distribution of Molybdenum Target X Radiation

Fa-jun CHEN,Si-chun ZHOU,Jin-jie WU,Wei JIANG,Bin GUO
DOI: https://doi.org/10.3969/j.issn.0258-0934.2014.11.002
2014-01-01
Abstract:The two-dimensional ion chamber matrix method and thimble ionization chamber scanning method are respectively used to measure the radiation field of X rays.The former can quickly measure radiation field, realize X-ray primary beam, aperture, shutter, filter, positioning device coaxial adjustment.The later is more accurate and reliable because of its consistency in scanning the uniform area of the radiation field point by point.When it is 1m away from the molybdenum target spot, the thimble ionization chamber can have a hori-zontal direction uniformity better than 95%with the radiation field size of about 124 mm,a vertical uniformity better than 95%with the radiation field size of about 103 mm, meeting the molybdenum target X-ray bench-mark establishment and quantity transmission requirements.
What problem does this paper attempt to address?