Tmente Carlo Simulation Method Application on Dosimetry Evaluating for Positioning Sensitivity of 2D Ionization Chamber Array

任民,吴昊,蒋幡,黄斐增,肖桂平,包尚联
DOI: https://doi.org/10.3969/j.issn.1005-202X.2010.04.002
2010-01-01
Abstract:Objective: Different kinds of 2D ionization chamber arrays may have different positioning sensitivities,advantages and shortcomings due to their distinct structure of element size,We investigated this problem using Monte Carlo(MC) simulation method to calculate dose distribution and evaluate their positioning sensitivities.Methods: The fluence distribution modeled for a 6 MV accelerator LINAC using BEAMnrc software and phase space file was the output of MC simulation.Dose distribution corresponded to a special leaf of multi-leaf collimator(MLC) location wasw calculated and evaluated for detector positioning sensitivity for both PTW Seven29 cube detector and IBA MatriXX cylinder detector respectively.Conclusion: The difference of positioning sensitivity is consistent in uncertainties of 2%,which is acceptable in the accuracy and precision QA/QC procession in Radiotherapy.