Calibration-free Sensor for hydrogen sulfide detection based on tunable laser absorption spectroscopy technique

Yan-jun ZHU,Xiao-hua FAN,Chao GAO,Jiang-long GUO
DOI: https://doi.org/10.14016/j.cnki.jgzz.2017.05.040
2017-01-01
Abstract:In order to provide a fast,accurate and sensitive hydrogen sulfide (H2S) detection device,the paper built a sensor based on tunable diode laser absorption spectroscopy (TDLAS) through a distributed feedback laser (DFB) as optical source.The laser wavelength scanning range covers the strongest absorption band of H2S.H2S absorption line at 1578.128nm is selected as target detection spectral line.A digital lock-in amplifier is used to demodulate the first harmonic (1 f) and the second harmonic signal (2f) signals simultaneously.Normalization of the 2f signal by 1f signal enables the sensor for calibration-free measurement so that the interference from light path,circuit and vibration can be restrained.Mass Flow Controller is used to configure 0-100ppm (parts-per-million) H2S standard gas for validating sensor performance.The results show the sensor has good linearity and stability,and the response time is about 20s.Moreover,the detection limit achievement by Allan variance is 160ppb(parts-per-billion) with an integration time of 30s.The sensitive H2S sensor can be applied to environmental monitoring in opening light path,and it is meaningful for real-time detection in many other fields.
What problem does this paper attempt to address?