Simultaneous detection of hydrogen sulfide and carbon dioxide based on off‐axis integrated cavity output spectroscopy using a near‐infrared distributed feedback diode laser

Xing Tian,Gang Cheng,Yanan Cao,Jiajin Chen,Kun Liu,Xiaoming Gao
DOI: https://doi.org/10.1002/mop.32887
IF: 1.311
2021-05-17
Microwave and Optical Technology Letters
Abstract:<p>An optical scheme for simultaneous detection of hydrogen sulfide (H<sub>2</sub>S) and carbon dioxide (CO<sub>2</sub>) is demonstrated. A near infrared tunable diode laser operating at 1.578 μm including the absorption lines of H<sub>2</sub>S at 6336.62 cm<sup>−1</sup> and CO<sub>2</sub> at 6336.24 cm<sup>−1</sup> was used as the light source for spectroscopic measurements of H<sub>2</sub>S and CO<sub>2</sub>. A high‐finesse cavity with a base length of 21 cm and a pair of high reflectivity mirrors with a reflectivity of 99.999% were employed based on off‐axis integrated cavity output spectroscopy. Allan variance analysis shows that the minimum detection limits of 150 ppbv for H<sub>2</sub>S in 170 s averaging time and 1.03 ppmv for CO<sub>2</sub> in 209 s, respectively.</p>
engineering, electrical & electronic,optics
What problem does this paper attempt to address?