Virtual Point Source Efficiency Calibration Technology Based on Virtual Point Detector

Tian Zining,Liu Wenbiao,Chen Wei,Long Bin,Feng Tiancheng,Wang Xuemei,Ouyang Xiaoping
DOI: https://doi.org/10.11884/hplpb201830.170278
2018-01-01
High Power Laser and Particle Beams
Abstract:The detection efficiency of?75 mm×25 mm of 241Am standard samples were studied by using laboratory HPGe detection.When we used detectors with the same type of structure to measure a given volume source,the linearity function be-tween the virtual point source and the virtual point detector was established by using LabSOCS software and Monte Carlo method.For a detector,the position of virtual point detector can be fixed by a point source placed at different height,and then,the posi-tion of virtual point source can be fixed by using their linearity function.The real calibration process will be carried out by placing a point source at its virtual position,namely,the volume source is replaced with the point source.In conclusion,the position of virtual point source can be obtained through the virtual point detector.This is a completely new technology method,which will be widely applied in radioactive measurement.