The Relative Detection Efficiency Calibration of CCD Detector
LIU Xiao-yan,YANG Yan-ji,ZHU Yue,CHANG Zhi,XIAO Jun,LU Jing-bin,WANG Yu-sa,CUI Wei-wei,YAO Ke,FU Yun-qing,CHEN Tian-xiang,WANG Juan,HUO Jia,HAN Da-wei,LI Wei,HU Wei,LI Mao-shun,LU Bo,ZHANG Yi,ZHANG Zi-liang,XUE Rong-feng,LIU Miao,CHI Lin-xiang,MA Ke-yan,CHEN Yong
DOI: https://doi.org/10.3969/j.issn.0258-0934.2016.02.007
2016-01-01
Abstract:Detection efficiency is a critical parameter for characterizing the performance of detectors , which needs to be calibrated carefully .The relative detection efficiency of CCD detector in Low Energy X -ray Tele-scope is calibrated by EBIT ( Electron Beam Ion Trap ) equipment in FUDAN University , where a HPGe ( high-purity germanium ) detector can be used as a standard detector .The relative efficiencies at energy corresponding to 6.7 keV、8.3 keV、9.1 keV、9.7 keV、10.4 keV、11.9 keV and 13.3 keV are measured in this test .The cali-bration method used in this paper is significant for the future detection efficiency calibration .