Dynamic Calibration Method for Detection Efficiency of Radioactive Source

Tian Zining,Ouyang Xiaoping,Han Bin,Chen Wei,Su Chuanying,Liu Wenbiao,Tian Yanjie,Feng Tiancheng
DOI: https://doi.org/10.11884/hplpb201628.151146
2016-01-01
Abstract:It is difficult to produce the radioactive standard gas source such as 1 3 3 Xe,which makes the calibration of gspec-trometers difficult.The uniform falling velocities of the 1 3 3 Ba disk source and 1 3 7 Cs point source were used to calibrate the detec-tion efficiency of 1 3 3 Xe volume source and 1 3 7 Cs linear source.The non-uniform falling velocities of 1 3 3 Ba disk source and 1 3 7 Cs point source were used to calibrate the detection efficiency of heterogeneous 1 3 3 Xe volume source and heterogeneous 1 3 7 Cs linear source.The results show that the detection efficiencies of 1 3 3 Xe were 0.1 55 (for ?46.8 mm × 20 mm sample)and 0.143 (for?63.5 mm×20 mm sample)respectively,obtained for the uniform falling velocity of the source,and the uncertainty was 8.4%(k =2).The deviation between the efficiency of the heterogeneously distributed volume source and the efficiency of uniform vol-ume source in 14%,and 23% for linear sources.Therefore,the deviation caused by heterogeneity must be corrected.
What problem does this paper attempt to address?