A Low-Power Stable Wideband Current Source for Acupuncture Point Skin Impedance Measurements

Changpei Qiu,Tianxia Zhao,Qiuping Li,Xin'an Wang,Kanglin Xiao,Bo Wang
DOI: https://doi.org/10.1155/2021/6650651
2021-01-01
Journal of Healthcare Engineering
Abstract:A low-power stable wideband current source for acupuncture point skin impedance measurements has been designed employing a differential architecture and negative feedback. The circuits extend bandwidth to 1 MHz, reducing harmonic distortion to 0.24% at 1 MHz. The output impedance is 37 MΩ at 100 kHz and 11 MΩ at 1 MHz. The stability of the output current of the current source when connected to different loads is below 0.1% at frequencies up to 500 kHz and increases to 0.74% at 1 MHz. The circuit was manufactured in a 0.13-μm CMOS technology and measured results are presented. The area of the current source is 0.09 mm2 and its consumption is 1.2 mW. It is intended for low-power acupuncture point skin impedance measurements.
What problem does this paper attempt to address?