Development and Application of High Resolution Photoionization Time-of-Flight Mass Spectrometer for Online Monitoring of Catalytic Processes

Li Qing-Yun,Hua Lei,Jiang Ji-Chun,Hou Ke-Yong,Qi Ya-Chen,Tian Di,Wang Hai-Yan,Li Hai-Yang
DOI: https://doi.org/10.11895/j.issn.0253-3820.150454
IF: 1.193
2015-01-01
Chinese Journal of Analytical Chemistry
Abstract:Methanol to olefins (MTO) technology is developed to produce fundamental organic chemical raw materials, the low carbon olefins, from coal and nature gas, which are firstly converted to methanol. MTO has become the key technique in coal to olefins route, and thus it is of great economic value and strategic significance. However, it is a great challenge for traditional analytical methods to real time and online analyze the catalytic products during MTO catalytic reaction process, which makes it difficult to optimize the catalytic reaction condition and explore the reaction mechanism. In this paper, in order to realize on. line monitoring of MTO catalytic reaction process, a home. made high resolution photoionization time. of. flight mass spectrometer (TOFMS) based on a vacuum ultraviolet (VUV) lamp was developed, using single photon ionization as the soft ionization source. The field. free region was 348. 5 mm, and the overall size of the TOFMS was 750 mm. 460 mm. 413 mm. As a result, a mass resolution of 3960 was achieved at m/z 28, and the mass peaks of N-2 and C2H4 were baseline separated, which avoided the interference of the carrier gas N-2 to the identification of the product C2H4 during MTO catalytic process. Under the condition of ionization region pressure 7 Pa, the limits of detection down to 0.51, 0.052, and 0.042 mL/m(3) were obtained for ethylene, propylene and butylene within 5 s, respectively. The instrument was successfully applied to the real. time, fast, and on. line analysis of gas phase products during MTO catalytic process. The results indicated that this instrument had potential application value and broad application prospects in industrial process monitoring.
What problem does this paper attempt to address?