Development of an integrated high-voltage electron microscope–gas chromatograph–quadrupole mass spectrometer system for the operando analysis of catalytic gas reactions

Longshu Tang,Tetsuo Higuchi,Shigeo Arai,Hiromochi Tanaka,Shunsuke Muto
DOI: https://doi.org/10.1093/jmicro/dfae010
2024-02-27
Microscopy
Abstract:Abstract This paper describes the development of a gas chromatography–quadrupole mass spectrometry system attached to a differential-pumping-type environmental cell of the reaction science high-voltage electron microscopy instrument at Nagoya University to distinguish unambiguously between different gas species with the same mass-to-charge ratio. Several model experiments were used to verify the efficacy of the newly proposed system, confirming its ability to analye the atomic-level structural changes during heterogeneous catalysts and the associated gas-reaction kinetics simultaneously, providing new insights into operando measurements in the field of environmental transmission electron microscopy.
microscopy
What problem does this paper attempt to address?