Study of On-line Organic Pollutants Measurement with Chemical Ionization Mass Spectrometry

LI Jian-quan,SHEN Cheng-yin,WANG Yu-jie,WANG Hong-mei,HAN Hai-yan,ZHENG Pei-chao,XU Guo-hua,CHU Yan-nan
DOI: https://doi.org/10.3969/j.issn.1673-6141.2008.05.008
2008-01-01
Abstract:Parameters of homemade proton transfer reaction mass spectrometry(PTR-MS)for on-line mea- surements of volatile organic compounds(VOCs)restricted in"Integrated Emission Standard of Air Pollu- tants"(GB 16297-1996)were optimized and measured.The individual VOCs in pure N_2 were measured at E/N=144 Td by PTR-MS and the results show that the main ions are the protonated compound and few fragmentations exist.The sensitivity arid limitation of homemade PTR-MS was obtained to 11 VOCs.The efficiency of the catalytical converter in gas inlet was tested.The precision of the PTR-MS was discussed.The results show that the homemade PTR-MS is a simple and effective tool for real time monitoring the VOCs restricted in"Integrated Emission Standard of Air Pollutants".