System Optimization Study of Hybrid K-edge/XRF Densitometer for Uranium-Plutonium Solution Measurement
Yan Zhang,Chun-Qing Fu,Jun Qiu,Jin-Hui Qu,Yu-Chao Yang,Ren-Bo Wang,Bin Tang
DOI: https://doi.org/10.1016/j.apradiso.2024.111367
IF: 1.787
2024-05-26
Applied Radiation and Isotopes
Abstract:In this paper, a comprehensive hybrid K-edge/XRF densitometer (HKED) device model is constructed using MCNP simulation. After the modeling process, a systematic simulation study is conducted to analyze the physical parameters and material selection of KED and XRF. The simulation results reveal that the optimal parameters for the X-ray tube are an X-ray source voltage of 160 kV and a 1 mm Fe filter. The sample should be placed in a vial with an inner diameter of 1.4 cm and an outer diameter of 2 cm. For the KED technique, the determined main parameters are a 1.9cm Fe filter rod and an inner diameter of 0.08cm for the collimator. For the XRF technique, the determined main parameters are a 0.01cm Gd filter and an inner diameter of 0.3cm for the collimator, with a detector angle of 150°. After selecting appropriate parameters, the average calibration factor Δμ of the KED technique was found to be 3.301 cm 2 ·g -1 , with a relative standard deviation (RSD) of 3.36%. Additionally, the comparison between the simulated and calculated values of uranium concentration revealed a minimum measurement error of 0.4%. The minimum detection concentration of KED for uranium solutions is approximately 1 g/L. For plutonium solutions ranging from 0.5 to 20g/L, linear fitting of the K a1 net peak area and plutonium concentration showed a coefficient of determination (R 2 ) of 0.999. The detection limit of XRF for plutonium measurement was 2.33✕10 -4 g/L. The linear fitting coefficients ( R 2 ) of uranium concentration versus K-edge transmission rate and plutonium concentration versus K a1 net peak area for the hybrid technique in measuring uranium-plutonium mixed solutions are determined as 0.999 and 0.996, respectively, demonstrating the response relationship of the HKED device to uranium and plutonium under different concentrations.
radiology, nuclear medicine & medical imaging,nuclear science & technology,chemistry, inorganic & nuclear