A method for optimum PSA setting in the absence of a pure α or β emitter and its application in the determination of (237)Np/(233)Pa.

Xiao-gui Feng,Qian-ge He,Jian-chen Wang,Jing Chen
DOI: https://doi.org/10.1016/j.apradiso.2014.01.029
IF: 1.787
2014-01-01
Applied Radiation and Isotopes
Abstract:In the application of liquid scintillation counting (LSC), the α/β discrimination is carried out with the function of pulse shape analysis (PSA), which requires the setting of the optimum PSA level. The optimum PSA are usually determined by the generation of cross-over plots, whereby a pair of vials, one containing a pure α emitter and the other a pure β emitter, is counted. However, in some cases such as the determination of 237Np/233Pa, a pure α emitter or a pure β emitter is not available. Therefore, we have developed a new approach to set the optimum PSA by measuring the sample itself of mixed α/β emitters. The count rate of the sample in the α-multi-channel analyzer changes monotonically with the increase of the PSA, and there is always an inflection point which is related to the optimum PSA. By fitting the data near the inflection point with the function y=ax3+bx2+cx+d, we can obtain the optimum PSA as −b/(3a), which can be used to determine the radioactivity of 237Np/233Pa. The results obtained with this new approach were in good agreement with those obtained by HPGe γ spectrometry that was calibrated with an LSC sample of 237Np/233Pa under a radioactive secular equilibrium. The new approach is promising to be used in simultaneous determination of gross α and β emitters, especially in the absence of a pure α or β emitter.