Agreement between transpulmonary thermodilution and pulmonary artery catheter thermodilution cardiac output measurements in pigs

祝胜美,吴健,何慧梁,诸绍君
DOI: https://doi.org/10.3321/j.issn:0254-1416.2008.12.021
2008-01-01
Abstract:Objective To evaluate the agreement between transpulmonary thermodilution and pulmonary artery catheter thermodilution cardiac output(CO)measurements in pigs.Methods Thirteen immature pigs weighing 22-25 kg were studied.Tracheal intubation was performed and the animals were mechanically ventilated after induction of anesthesia.In each animal,some predetermined methods were used in turn to intervene in the hemodynamics.CO measurements were performed simultaneously by transpulmonary thermodilution(COTP)and pulmonary artery catheter thermodilution(COPA)immediately before intervention and at least 10 min following stable hemodynamics during each intervention.The agreement was assessed using Bland-Ahman analysis,and the interchangeable limits of the two methods were defined as the relative error of COTP less than 20%.Results CO measurements yielded a total of 205 data pairs with a range of 1.4-7.06 L/min for COTP and 1.25-6.55 L/min for COPA and the value of COTP was about 2% higher than that of COPA.The bias of COTP was 0.07 L/min and 95% confidence intenral(CI)was 0.03-0.11 L/min;the limit of agreement was-0.46-0.60 L/min and 95% CI for the lower and upper hmits of agreement was-0.53--0.40 L/min and 0.54-0.66 L/min respectively.The relative error of COTP was 15.3%.Conclusion There is a good agreement between transpulmonary thermodilution and pulmonary artery catheter thermodilution CO measurements in pigs.