Comparison of noninvasive cardiac output and stroke volume measurements using electrical impedance tomography with invasive methods in a swine model

Chi Ryang Chung,Ryoung Eun Ko,Geuk Young Jang,Kyounghun Lee,Gee Young Suh,Yongmin Kim,Eung Je Woo
DOI: https://doi.org/10.1038/s41598-024-53488-0
IF: 4.6
2024-02-06
Scientific Reports
Abstract:Pulmonary artery catheterization (PAC) has been used as a clinical standard for cardiac output (CO) measurements on humans. On animals, however, an ultrasonic flow sensor (UFS) placed around the ascending aorta or pulmonary artery can measure CO and stroke volume (SV) more accurately. The objective of this paper is to compare CO and SV measurements using a noninvasive electrical impedance tomography (EIT) device and three invasive devices using UFS, PAC-CCO (continuous CO) and arterial pressure-based CO (APCO). Thirty-two pigs were anesthetized and mechanically ventilated. A UFS was placed around the pulmonary artery through thoracotomy in 11 of them, while the EIT, PAC-CCO and APCO devices were used on all of them. Afterload and contractility were changed pharmacologically, while preload was changed through bleeding and injection of fluid or blood. Twenty-three pigs completed the experiment. Among 23, the UFS was used on 7 pigs around the pulmonary artery. The percentage error (PE) between CO UFS and CO EIT was 26.1%, and the 10-min concordance was 92.5%. Between SV UFS and SV EIT , the PE was 24.8%, and the 10-min concordance was 94.2%. On analyzing the data from all 23 pigs, the PE between time-delay-adjusted CO PAC-CCO and CO EIT was 34.6%, and the 10-min concordance was 81.1%. Our results suggest that the performance of the EIT device in measuring dynamic changes of CO and SV on mechanically-ventilated pigs under different cardiac preload, afterload and contractility conditions is at least comparable to that of the PAC-CCO device. Clinical studies are needed to evaluate the utility of the EIT device as a noninvasive hemodynamic monitoring tool.
multidisciplinary sciences
What problem does this paper attempt to address?
The paper aims to address the following issues: 1. **Comparison of non-invasive and invasive cardiac output (CO) and stroke volume (SV) measurement methods**: The main objective of the study is to compare the CO and SV measurements obtained using a non-invasive Electrical Impedance Tomography (EIT) device with those from three invasive devices (Ultrasonic Flow Sensor [UFS], Pulmonary Artery Catheter Continuous Cardiac Output [PAC-CCO], and Arterial Pressure Cardiac Output [APCO]) in a pig model. 2. **Evaluation of the performance of the EIT device**: Specifically, the study evaluates the performance of the EIT device in measuring dynamic changes in CO and SV in a mechanically ventilated pig model under different conditions of cardiac preload, afterload, and contractility, and compares it with invasive devices (particularly the PAC-CCO device). Through these comparisons, the paper aims to validate the potential clinical application of the EIT device as a non-invasive hemodynamic monitoring tool and provide a basis for further clinical research.