Uncertainty Evaluation of Catalytic Activity Concentration of GGT with Reference Measurement Procedure with Monte Carlo Method

WANG Hui-min,SHEN Lei,WANG Yan-qiu,JI Huo-yan,XIAO Jing,WANG Jian-xin,CHEN Long-mei
DOI: https://doi.org/10.3760/cma.j.issn.1009-9158.2013.09.014
2013-01-01
Abstract:Objective To establish the methodology of uncertainty evaluation with reference measurement procedure by Monte Carlo method (MCM).Methods According to JCGM 101:2008,we established the methodology of uncertainty evaluation by MCM in the example of GGT reference measurement procedure.We could calculate an estimate of GGT concentration,the associated standard uncertainty and a coverage interval with a specified probability by MATLAB software,setting the uncertainty evaluation by GUM method as a control.Results When the uncertainty was evaluated by GUM method,the results of sample A and Sample B were (95.8 ±2.4) U/L (k =2) with coverage interval [93.4 U/L,98.2 U/L] and (180.0 ± 3.9) U/L (k =2) with coverage interval [176.1 U/L,183.9 U/L] respectively,while using MCM method,the uncertainty evaluation result of sample A and Sample B were (95.8 ± 2.4) U/L (k =2) with coverage interval [93.4 U/L,98.2 U/L] and (180.0 ± 3.9) U/L with symmetrical 95% coverage interval [176.2 U/L,183.8 U/L].The output quantity simulated by MCM was normal distributed.Conclusions When the distribution of the output quantity is normal,the measurement uncertainty evaluated by both MCM and GUM method is nearly the same.When the distribution of the output quantity is unknown,MCM can be used as a verification of GUM method.
What problem does this paper attempt to address?