Application of Reference Procedure for the Measurement of Gamma - Glutamyl-Transferase in Clinical Biochemistry

WANG Jianbing,ZHUANG Junhua,ZHENG Songbai,ZHOU Jingjing,LIN Li,HUANG Xianzhang
DOI: https://doi.org/10.13602/j.cnki.jcls.2009.05.009
2009-01-01
Abstract:Objective To explore the application of reference procedure for the measurement of gamma-glutamyltransferase (GGT).Methods The routine assays were calibrated by the serum GGT levels,which were measured by the reference procedure.The GGT in three external quality materials,four commercially available samples,and five fresh sera was measured in parallel by the routine assays and the reference procedure.Results Before the routine assays were calibrated,the difference of GGT levels measured by each of routine assays 4,5,and 9 and by the reference procedure was higher than 10%.The bias of low value between routine system 8 and the reference procedure was exceeded 10%.The bias of GGT between each of routine assays 1,2,3,6,and 7 and the reference procedure was less than 10%.The percentage of the bias more than ±5% between each of the routine assays and the reference procedure was 38.9%,33.3%,and 4.4% in measuring GGT in the commercially available samples,external quality materials,and fresh sera respectively.Conclusions Fresh serum in which GGT activity was assigned by the reference procedure may be used to dissolve the issues of the traceability and comparability.