Measurements comparison of Sirius and Orbscan-Ⅱ on transverse diameter of corneal in myopia and anterior chamber depth

Lemei Qiu,Xiaolan Zhang,Shunqing Wang,Chengshu Sun,Yingping Deng
DOI: https://doi.org/10.3760/cma.j.issn.1006-4443.2014.011.006
2014-01-01
Abstract:Objective To compare Sirius and corneal measurement of Orbscan-Ⅱtransverse diameter and anterior chamber depth,and to provide the reference for clinical application.Methods In this prospective study,patients in the Department of Ophthalmology of West China Hospital of Sichuan University were included from November 2012 to February 2013,and with eye symptoms,corneal operation history and anterior segment pathologic changes were eliminated.Ninety-seven cases of myopia patients (97 eyes) were selected with corneal thickness averaged 527 m,the corneal transverse diameter (W-W) and anterior chamber depth (ACD) were measured using both Sirius and Orbscan-Ⅱ instrument before the excimer laser operation,the differences between two kinds of measurement were statistically compared by using paired t test and SPSS17.0 software,the regression equation were drawn.Results Average of corneal diameter (W-T-W) measured by Sirius and Orbscan-Ⅱ were (11.86 ± 0.38) mm and (11.50 ± 0.31) mm respectively,the difference between the average was (0.36 ± 0.23) mm,there was statistically significant (P =0.000 or P <0.05); Average of ACD was (3.28 ± 0.22) mm and (3.13 ± 0.24) mm respectively,the difference between the two averages were (0.14 ±0.18)mm (P =0.00,P <0.05),the difference was statistically significant.The differences of two instruments were statistically significant (P <0.05),the correlation coefficients were r1=0.785,P =0.00; r2=0.694,P =0.According to the regression equation,the W-W and ACD measured by Sirius were better than Orbscan test results.Sirius (W-W) =(5.155±0.535)× Orbscan-Ⅱ (W-W) mm,Sirius (ACD) =(0.538±0.790)× Orbscan-Ⅱ (ACD) mm.Conclusions Sirius and Orbscan-Ⅱ can accurately measure the normal eye corneal diameter and anterior chamber depth,the test results by two kinds of examination can be filled in the formula for exchange,and it provide the references for clinical.
What problem does this paper attempt to address?