Comparison of central corneal thickness measurements between anterior segment optical coherence tomography and ultrasonic pachymeter after DSAEK

Bing-zhen LI,Rong-mei PENG,Jing REN,Feng LIU,Ling-ling WU,Jing HONG
DOI: https://doi.org/10.3760/cma.j.issn.1006-4443.2011.07.014
2011-01-01
Abstract:Objective To assess the agreement of central corneal thickness ( CCT) measured by anterior segment-optical tomography (AS-OCT) and ultrasonic pachymeter after Descemet's Stripping with Automated Endothelium Keratoplasty (DSAEK) and provide objective basis for clinical application of AS-OCT. Methods Forty-one eyes of 38 patients after Descemet's Stripping with Automated Endothelium Keratoplasty (DSAEK) underwent CCT measurement by AS-OCT and ultrasonic pachymeter.The data was analyzed by paired-t test and Pearson correlation analysis. Bland-Altman plot was used to assess the agreement.Results The mean CCT values were(728.73±1 38.39)μ m measured by AS-OCT and (718.56+131.37)μ m measured by ultrasonic pachymeter,and there was no significant difference between the two methods. Regression analysis showed a high correlation between the values obtained by both devices (r =0.916, P<0.01). AS-OCT and ultrasonic pachymeter showed a less a-greernent in CCT measurement after DSAEK. Conclusions After DSAEK, there is a reasonable agreement between two techniques, AS-OCT and ultrasonic pachymeter of CCT measurements. Considering the special advantages, AS-OCT is an effective method to observe the changes of the corneal thickness in the long term.