Rapid Determination and Continuous Monitoring of Propofol in Microliter Whole Blood Sample During Anesthesia by Paper Spray Ionization-Mass Spectrometry

Liu Ying,Zhang Xiao-Hui,Mi Wei-Dong,Zhou Ying-Lin,Zhang Chang-Sheng,Zhang Xin-Xiang
DOI: https://doi.org/10.1007/s00216-020-02999-6
2020-01-01
Analytical and Bioanalytical Chemistry
Abstract:Propofol is a widely used intravenous anesthetic agent in sedation and general anesthesia. To improve the safety and maintain the depth of anesthesia, it is important to develop a rapid, sensitive, and reliable method to monitor the concentration of propofol in blood during anesthesia continuously. Here, we present a novel strategy based on paper spray ionization-mass spectrometry (PSI-MS) to detect propofol. Samples (in 10 μL) were mixed with methanol as protein precipitation solvent and 2,6-dimethylphenol as internal standard. Protein micro-precipitation was achieved with methanol by vortexing and centrifuging for 5 s each, and propofol was extracted to the supernatant. PSI-MS was performed in negative ionization mode, and MS signal lasted for 1 min. The analysis of a single sample was completed within 2 min. The area ratios of propofol to internal standard were calculated for quantification. Limit of detection of 5.5 ng mL−1 and limit of quantification of 18.2 ng mL−1 were achieved for propofol in whole blood. Calibration curve was linear in the range of 0.02–10 μg mL−1. The developed method was used successfully in monitoring the propofol concentration in 3 patients’ whole blood during anesthesia, showing its further application in controlling and feeding-back target concentration infusion.
What problem does this paper attempt to address?