Simultaneous On-Line Monitoring of Propofol and Sevoflurane in Balanced Anesthesia by Direct Resistive Heating Gas Chromatography.

Hao Dong,Feng Jiang Zhang,Fu Yuan Wang,Ying,Jing Guo,G. M. Kanhar,Jing Chen,Jun Liu,Chen Zhou,Min Yan,Xing Chen
DOI: https://doi.org/10.1016/j.chroma.2017.05.001
IF: 4.601
2017-01-01
Journal of Chromatography A
Abstract:In balanced anesthesia, sevoflurane and propofol are often used in combination to achieve a better anesthetic effect. However, methods for on-line monitoring of concentrations of the two anesthetics in patients are still rare in clinical. This study proposed a non-invasive method utilizing a fast gas chromatograph combined with a surface acoustic wave sensor (Fast GC-SAW) to simultaneously on-line monitor sevoflurane and propofol in patients' exhaled gas. By using the direct resistive heating capillary column, the single detection time of Fast GC-SAW system was significantly shortened to 90s, as well as the size reduced to (40cm x 30cm x 20 cm). Besides, in the calibration of sevoflurane, Fast GC-SAW system showed a good linear correlation (R-2 = 0.9925, P < 0.01) with gas chromatography-mass spectrometer (GC-MS), which ensured the reliability and accuracy of the Fast GC-SAW system. Finally, clinical exper. iments on patients under balanced anesthesia were conducted. The varied concentrations measured by Fast GC-SAW extraordinarily matched the clinical usages of these two anesthetics. (C) 2017 Elsevier B.V. All rights reserved.
What problem does this paper attempt to address?