Optimization PEM designs based on Monte Carlo simulations

Mengdie Wang,Yilong Liu,Guangshu Hu,Hui Zhang
DOI: https://doi.org/10.16511/j.cnki.qhdxxb.2014.08.020
2014-01-01
Abstract:The influence of the detector thickness in dual-plate positron emission mammography (PEM) on the reconstructed image quality was evaluated to optimize the crystal thickness.The evaluations used a dual-plate PET system modd built on the GEANT4 Application for Tomographic Emission (GATE) platform.Monte Carlo simulations were conducted for various detector thicknesses to analyze the system performances.Comparisons of the system sensitivity,image spatial resolution and tumor detectability of different systems showed that thinner crystals tended to improve the spatial resolution but reduced the system sensitivity and tumor detectability.The spatial resolution along the two directions of the panel's width and length were both good for a 10 mm thick crystal and were not severely impacted by the depth of interaction (DOI) effect.The system sensitivity was about 11%.With 10 min data acquisition,a small low concentration tumor source could be easily distinguished from the background in the reconstructed image.Consequently,10 mm was chosen as a optimal crystal thickness for the PEM system to achieve both higher system sensitivity and less DOI effect.