Pressure Measurement and Compensation for Mid-infrared Methane Detection

Liu Zhi-wei,Li Zi-wen,Li Ya-fei,Zheng Wen-xue,Zheng Chuan-tao,Wang Yi-ding
DOI: https://doi.org/10.3788/gzxb20184702.0230002
IF: 0.6
2018-01-01
ACTA PHOTONICA SINICA
Abstract:A pressure measurement and compensation technique was studied by employing a 3.291 mu m Continuous Wave ( CW) Interband Cascade Laser (ICL) and a dense-patterned Multipass Gas Cell (MPGC) with an effective optical path length of 54.6 m. The pressure inside the MPGC was measured based on direct Lorentzian absorption line fitting on the measured absorption spectral signal of CH4 , and then pressure compensation was made on the masured CH4 concentration. Pressure calibration was performed from 1.33 x 10(4) Pa to 10.64 X 10(4) Pa using a 2.1 X 10 (6) CH4 sample. An Allan deviation analysis of the measured pressure of a 2.1x10 (6) CH4 at 9.31 X 10(4) Pa pressure indicates a measurement precision of similar to 219.5 Pa with a 2. 2 s averaging time. Fiveteen groups of pressure/concentration measurements of 1.0 x 10 (6) , 1.2x10 (6) , 1.4x10 (6) , 1.6x10 (6) and 2.1x10 (6) CH4 samples at different pressures of 1.33 x 10(4) , 3.99 x 10(4) and 6.65 x 10(4) Pa were performed, and the results proved the feasibility of the proposed pressure measurement and compensation technique.
What problem does this paper attempt to address?