CH4/C2H6 dual gas sensing system using a single mid-infrared laser

Jiachen Sun,Jun Chang,Yu Zhang,Yubin Wei,Qinduan Zhang,Fupeng Wang,Shan Lin,Zihan Wang,Minghui Mao
DOI: https://doi.org/10.1016/j.saa.2023.122368
2023-04-01
Abstract:Methane (CH<sub>4</sub>) and ethane (C<sub>2</sub>H<sub>6</sub>) dual gas sensor with low system complexity and strong stability is proposed. The correction method based on absorbance spectrum is applied, and the cross-interference of C<sub>2</sub>H<sub>6</sub> to CH<sub>4</sub> is eliminated. In the single gas concentration measurement, linear fitting is performed between the absorbance and concentration of CH<sub>4</sub> and C<sub>2</sub>H<sub>6</sub>, and the correlation coefficients of R<sup>2</sup> = 0.99959 and R<sup>2</sup> = 0.99994 are obtained respectively, which proves that the accuracy of the dual gas sensor is robust. In the dual gas concentration measurement, we carry out continuous measurement of five mixed gases and a long-term measurement of a mixture of gases, which verifies that our sensor has the fast response speed and strong stability. The minimum detectable column densities of 0.62 ppm·m for CH<sub>4</sub> and 0.1 ppm·m for C<sub>2</sub>H<sub>6</sub> are achieved, respectively. The CH<sub>4</sub>/C<sub>2</sub>H<sub>6</sub> dual gas sensor assisted by the correction method has high sensitivity and strong robustness to cross-interference, and has great potential for application in various scenarios.
spectroscopy
What problem does this paper attempt to address?