A High-Balance Planar SQUID Gradiometer

Xue Zhang,Guofeng Zhang,Yongliang Wang,Liangliang Rong,Shulin Zhang,Jun Wu,Longqing Qiu,Xiaoming Xie,Zhen Wang
DOI: https://doi.org/10.1109/tasc.2019.2898678
IF: 1.9489
2019-01-01
IEEE Transactions on Applied Superconductivity
Abstract:We have developed and fabricated SQUID gradiometers in a Nb/Al-AlOx/Nb technology using an i-line stepper combined with a mask aligner lithography. The stepper and the aligner are used to fabricate fine-scale patterns such as the Josephson junctions and large-scale patterns such as pickup coil, respectively, on the same wafer with a 4-inch diameter. The SQUID is designed as a first-order gradiometer with a dual-washer configuration. Two sets of multiturn spiral input coils with opposite directions are integrated on the washer and connected to the pickup coil to form a flux transformer. In order to improve the balance of the gradiometer, a fully symmetric configuration is introduced with a mirrored SQUID. The pickup coil is a first-order gradiometric type with a base line of 3.5 cm. The size of each loop is 12.65 mm X 11.65 mm, which results in a gradient magnetic field sensitivity of 0.25 nT/(cm.Phi(0)). The measured gradient magnetic field noise is 75 fT/(m.root Hz) in the white noise region. Moreover, a sapphire holder is used to fix and package the chip in order to maintain the high intrinsic balance (typical 10(-4)). It makes the SQUID gradiometer more attractive for an unshielded operation in noisy environments.
What problem does this paper attempt to address?