Biomechanical effect of osteoporosis on sacral pedicle screws instrumentation

YU Binsheng,ZHENG Zhaomin,ZHUANG Xinming
DOI: https://doi.org/10.3969/j.issn.1004-406X.2009.10.07
2009-01-01
Abstract:Objective:To investigate the biomechanical effect of osteoporosis on sacral pedicle screws instru-mentation.Method:25 fresh adult human sacra specimen featuring osteoporosis were used in this study.Accord-ing to the value of lumbar bone mineral density(BMD) assessed by a dual energy X-ray absorptiometry, specimens were assigned into three groups as follows,group A(n=9),BMD 0.70-0.79g/cm2,group B(n=8),BMD 0.60-0.69g/cm2,group C(n=8),BMD<0.60g/cm2.For each specimen,S1 pedicle screw was inserted bicortically on the left side,and S1 pedicle screw augmented by polymethylmethacrylate(PMMA) was inserted unicortically on the contralateral side of sacrum.Following a dynamic cyclic loading from 30N to 250N on the screw head for 2000 cycles,the subsidence displacement and axial pull-out strength of each screw were measured,and we defined subsidence displacement over 2mm after 2000 cyclic loading as anchoring failure.Result:No anchoring failure was noted in both group A and B.However,in group C,6 cases(75%) in bicortical fixation and 5 cas-es(63%) in PMMA augmented fixation had evidence of failure during cyclic loading.In group A,no significant difference between bicortical and PMMA augmented fixation was detected in terms of the subsidence displace-ment and maximal screw pull-out strength(P>0.05).In group B,significantly less subsidence and higher maxi-mal screw pull-out strength were documented in the PMMA augmentation than that in bicortical fixation(P< 0.05).Both techniques in group A exhibited lower subsidence of the screw than that in group B(P<0.05).Bi-cortical technique in group A exhibited higher maximum pull-out strength than that in group B(P<0.05).How-ever,statistical difference in terms of PMMA augmentation was not detected between group A and B(P>0.05).Among the cases with anchoring failure in group C,significantly more loading cycles(before failure) were ob-served in PMMA augmented unicortical fixation than that in bicortical fixation .Conclusion:For BMD ≥ 0.70g/cm2,bicortical sacral pedicle screw fixation could obtain sufficient anchoring strength compared to that provide by PMMA augmented technique.When BMD is 0.60-0.69g/cm2,PMMA augmented technique have more pull-out screw strength than bicortical fixation.For BMD<0.60g/cm2,early screw loosening may occur in both bicortical and PMMA augmented fixations.Therefore,preoperative evaluation of lumbar BMD is critical to choose a appropriate technique for sacral screw instrument.
What problem does this paper attempt to address?