Evaluation of Automated Quantitative D-Dimer Test on the VIDAS Instrument

周静,王霞,廖娟,江虹
DOI: https://doi.org/10.3969/j.issn.1671-7414.2008.05.015
2008-01-01
Abstract:Objective To evaluate the performance feature of the automated quantitative D-dimer test on the VIDAS instrument.Methods According to the standard of the NCCLS documents,investigated the quality of the quantitative D-dimer test on the VIDAS instrument.Results Intra-assay precision of high level control,low level control and mixed plasma was 3.11%,3.43% and 6.17% respectively.Extra-assay precision of high level control,low level control and mixed plasma was 4.35%,4.10% and 7.48% respectively.Total reproducibility was 8.2%,recovery was 102%.The measurement range from 45 ng/ml to 4 641.50 ng/ml was evaluated.It showed good linearity during this range.The influence of hemoglobin(0.3 mmol/L),triglyeride(2 mmol/L),bilirubin(125 μmol/L),rheumatoid factor(200 IU/ml)were less than the given coefficient of variation(10%).20 tested healthy plasma validated the given reference range was used in their lab.Conclusion Quantitative D-dimer test on the VIDAS instrument has good accuracy,precision and linearity.Its convenience,automation is suitable for individual test in emergency.