Performance verification of diluted thrombin time assay for detecting Dabigatran level in plasma

Xue-lian WU,Chen-xue QU,Ju-hua DAI,Li-ping LI,Yan GONG,Yao LU,Jia-ying YUAN,Lian-fang NI
DOI: https://doi.org/10.13602/j.cnki.jcls.2017.12.07
2017-01-01
Abstract:Objective To evaluate the performance of diluted thrombin time (dTT) assay for detecting Dabigatran levels and observe whether this assay may meet the requirements of clinical laboratory.Methods According to EP15-A2,EP6-A,EP7-A and C-24 documents of the Clinical and Laboratory Standards Institute (CLSI),the precision,trueness,analytical measurement range,carryover rate and anti-biological interference of dTT assay were evaluated and the stability of specimen for dTT assay was observed.Results Both the within-day and between-day coefficient of variation (CV) of dTT assay for detecting Dabigatran levels were consistent with manufacturer's stated CV.Compared with target values of Dabigatran,the relative bias of 3 levels of proficiency test materials from College of American Pathologists (CAP) were less than 10%.The results meet linear verification when Dabigatran concentration was between 30.92 and 249.13 ng/mL.The carryover rate was-0.84%.There was no interference for Dabigatran levels by dTT assay for detecting Dabigatran when Hb≤3 g/L,triglyceride≤873 mg/dL,heparin≤2.2 IU/mL and FDP≤29 mg/L.The results of stability showed that plasma specimens for dTT could not be stored at room temperature more than 4 hours,at 4 ℃ more than 4 days,at-20 ℃ exceed 1 month,while at-80℃ the plasma specimens could be stored at least 6 months for dTT assay.Conclusion The precision,trueness,analytical measurement range,carryover rate,anti-biological interference of dTT assay may meet the requirement of clinical laboratory.The stability of the specimen can fulfill the clinical requirements.
What problem does this paper attempt to address?