Application of Innovance D-dimer in Clinical Settings with Cutoff Evaluated

TONG Xue-dong,WANG Cheng-bin,LI Jian
DOI: https://doi.org/10.3969/j.issn.1009-6213.2012.03.011
2012-01-01
Abstract:Objective To investigate performance of Innovance D-Dimer measuring clinical samples on CS2000 automatic coagulation analyzer with its original cutoff.Method Samples from 100 patients with different diseases were test by STA Liatest D-Dimer(with STA Evolution) and Innovance D-dimer respectively,bias between two sets of data and their correlation were analyzed,correspondent rate were calculated according cutoff ruling out venous thromboembolism.Results Data from two methods are correlated statistically not only in analysis range(r=0.9304,P<0.05),but also in undiluted sample with D-dimer lower than 4 mg/L(r=0.9304,P<0.05).Negative,positive and whole correspondent rate were 92.3%、98.6% and 97% taking 0.5 mg/L as cutoff.Conclusion Innovance D-dimer is comparative and correspondent with main stage method in clinical application.