Structural Deviation Calibration and Parameters Obtaining Algorithm for Linear Array Detector CT System

YAO Ming,CONG Peng,LIU Xi-ming
DOI: https://doi.org/10.3969/j.issn.0258-0934.2013.01.001
IF: 1.279
2013-01-01
Nuclear Instruments and Methods in Physics Research Section B Beam Interactions with Materials and Atoms
Abstract:Deviation of geometric parameter from its design value of linear array detector computed tomography(CT) would result in image quality degradation.A structural deviation calibration and parameters obtaining algorithm based on projections is developed to improve the accuracy.The algorithm uses a simple calibration phantom and serial projections to adjust the system structure and analytically estimate the geometric CT parameters.Calculations demonstrate that the algorithm is very accurate with the error less than 0.5%,and experiments show high quality images.