CT geometrical deformation calibration method in animal dual-modality CT-SPECT system

JI Qi-fan,LI Jun,CAO Wen-tian,BAO Shang-lian
DOI: https://doi.org/10.13929/j.1003-3289.2010.11.062
2010-01-01
Abstract:Objective: To propose a new method, in order to correct the geometrical deformation existed in mechanism system in a flat-panel detector based on cone beam CT and pinhole SPECT assembled dual modality small animal CT-SPECT system developed in our laboratory. Methods: A Cartesian coordinate system was defined for the dual modality small animal imaging system, the Powell algorithm was used to address the geometrical parameters. Results: In the Cartesian coordinate system, origin was fixed on the center of flat-panel detector, and all geometrical parameters were defined and calculated. After the corrections, geometrical artifacts were removed effectively, and the quality of the image was greatly improved. Conclusion: This method can accurately correct the geometrical deformations in dual modality small animal CT-SPECT system. This approach is also applicable to other cone-beam X-ray computed tomography imaging system, and it is also a solid base for SPECT geometrical correction, as well as the image registration between CT and SPECT images.
What problem does this paper attempt to address?