Projection angle based geometric correction for micro cone-beam CT

CAO Wen-tian,LI Cheng-chao,LI Jun,BAO Shang-lian
DOI: https://doi.org/10.13929/j.1003-3289.2011.11.010
2011-01-01
Abstract:Objective To improve the spatial resolution of cone-beam CT(CBCT) for small animal.Methods Firstly,360° scan of a three-ball phantom was performed.Then,7 geometric parameters such as the position of the X-ray source,shift of the axis of rotation,etc.The projection of the center of the balls were calculated by centroid estimation.Finally,the geometric parameters were determined at each angle for the instability of the system during the rotation.Image reconstruction was done by FDK algorithm with projection angle depended on geometric parameters.Results The resolution of the reconstructed image was improved for about 2 lp/mm.Conclusion Projection angle based geometric correction can solve the problem of small variance of the geometric parameters during scanning,and significantly improve the spatial resolution of CBCT.