Projection Weighting Reconstruction Method for CBCT System with an Off-Center Detector

郝佳,李亮,张丽,康克军
DOI: https://doi.org/10.16511/j.cnki.qhdxxb.2013.02.006
2013-01-01
Abstract:CBCT (cone-beam CT) scanners have been widely used in medical diagnosis and local tissue imaging because of low-dose and flexibility. To reduce the hardware cost and radiation dose, CBCT scanners always use off-center detector scanning geometry, which can keep a reasonable field of view (FOV) and use a smaller size detector. However, the projections are transversely truncated under this scanning condition, and reconstruction using conventional filtered back-projection reconstruction methods will lead to serious truncation artifacts and errors. A reconstruction method was developed based on projection pre-weighting to keep the corrected projections smooth and continuous and the back-projection coefficient consistent. Numerical simulation was conducted using FORBILD jaw phantom with different scanning geometries and detector truncation conditions. CBCT projections were acquired and then reconstructed using the proposed method. The experimental results show that when the detector size is transversely reduced less than 30%, this method can acquire similar and excellent results compared with the FDK reconstruction results from a full-size detector. The method is easily implemented and highly efficient, which is feasible in an off-center CBCT system.
What problem does this paper attempt to address?