HPLC Determination with Fluorescent Detection of Telmisartan Concentration in Plasma and Its Application

况荣华,张会杰,熊玉卿
2005-01-01
Abstract:Objective:To develop an HPLC method with fluorescent detection for the determination of telmisartan concentration in human plasma and apply to pharmacokinetic study of teimisartan.Methods:Telmisartan in plasma was extracted with solid phase extraction method.The column was Diamonsil C_(18)(5μm,4.6mm×150mm).The mobile phase consisted of acetonitrile-potassium dihydrogen phosphate buffer solution(61:39,adjusted pH to 3.74 with H_3PO_4)at a flow rate of 1 mL·min~(-1).The internal standard was candesartan.The fluorescence detector was operated at λ_(Ex)=305 nm and λ_(Em)=365nm.Results:The calibration curve of telmisartan was linear in the range of 0.5-144 ng·mL~(-1)(r=0.9998).The absolute recovery was 79.62%-85.69%(n=5),the relative recovery was 101.9%-109.0%(n=5),and the intra-day and inter-day RSD were 4.23%-9.80% and 4.03%-9.95%(n=5)respectively.Conclusion:This method possesses merits of good sensitivity,accuracy and simplicity.It is shown to be suitable for clinical monitoring concentration of telmisartan in human plasma and phar- maeokinetic studies of telmisartan in humans.