Evaluation of HPLC -Fluorescence and LC -ESI/MS for Determination of Telmisartan in Human Plasma

SHEN Yu-lan,FENG Fang,WU Jie
2006-01-01
Abstract:Objective:To establish an HPLC-fluorescence method for the determination of telmisartan in human plasma,and we also compared the differences between HPLC-fluorescence and LC-ESI/MS.Method:Mirtazap- ine was used as internal standard in the method of HPLC-fluorescence.After adding mirtazapine,the plasma sam- ples were extracted with ethyl acetate and determined by HPLC-fluorescence.The separation was carried out on a ODS column(150 mm×4.6 mm,packed with 5μm C_(18)silica RP particle)at 35℃.The mobile phase was methanol -0.01 mol·L~(-1)monobasic potassium phosphate(pH=6)(62:38,v/v),running at a flow rate of 1.0 mL· min~(-1).In the method of HPLC-MS,the plasma samples were extracted with ethyl acetate and performaneed in se- lected ion monitoring(SIM)mode.Result:The linear calibration curves were both obtained in the concentration range of 0.5-400 ng·mL~(-1)in HPLC-fluorescence method and HPLC-MS method.The detection limit of telm- isartan in plasma in the method of HPLC-fluorescence was 0.5 ng·mL~(-1)( the volume of plasma is 0.2 mL),and was 0.1 ng·mL~(-1)(the volume of plasma is 0.5 mL)in the method of HPLC-MS.The average recovery of two methods were both more than 80%.Conclusion:No significant difference was found in the results measured be- tween two methods through t-test.The sensitivity and accuracy of the HPLC-fluorescence is as good as HPLC- MS,while HPLC-fluorescence is simple and cheap.