Dosimetric Impact of Patients in Three Dimensional Conformal Radiotherapy of Esophageal Carcinoma by Using Matching Function of Simulator

汪志,唐虹,曹泓历,李锐,洪浩
DOI: https://doi.org/10.3969/j.issn.1006-6586.2010.03.010
2010-01-01
Abstract:Objective: To analyze changes of setup errors before and after using matching function of simulator for esophageal carcinoma, then discuss dosimetric changes caused by setup errors. Methods: Setup errors of 18 patients with esophageal carcinoma in the left-right, superior-inferior and anterior-posterior directions were obtained by using function of Varian Acuity8.6 simulator, and patients’ position was adjusted according to setup errors. After adjustment new setup errors were recorded, then their corresponding plans(plan2, plan3) were simulated in the Treatment Plan System. Meanwhile, they were compared to original plan (plan1). Results: The mean dose of PTV, CTV, GTV in plan3 was higher than plan2, but was lower than plan1 due to setup errors. In plan1, no patient had maximum dose of spinal cord over 4500cGy; While in plan2 and plan3 there were 2 patients had the spinal cord dose more than 4500cGy. In plan2, the heart V50, the mean dose of heart and lungs was highest and the lung V20 was lowest. Conclusions: The mean dose of target volume decreased and dose of Organs At Risk (OARs) increased due to setup errors. After matching the mean dose of target volume increased and the dose of partial OARs decreased than before matching, but the maximum dose of spinal cord and lungs V20 increased.