The Investigation of the Characteristics of a New Thimble Chamber

SUN Wen-zhao,CHEN Li-xin,SUN Hong-qiang,HUANG Shao-min,DENG Xiao-wu
2009-01-01
Abstract:Objective: To do comprehensive quality assurance for radiation measurement equipment is a basic component of clinic quality assurance program. This work investigated the characteristics of the new thimble chamber Exradin A19 in the measurement of electron beam and photon beam. Method: The ND factor of new chamber was calibrated under gamma ray of cobalt-60,the leakage current, reproducibility, angular dependence, stem effect, energy response, recombination effect and po- larity effect were evaluated for electron beam and Cobalt-60 gamma ray. Result: The reproducibility of the 11 readings is less than 0.05%;for Cobalt-60 gamma ray, the angular dependence of perpendicular to chamber axis is less than 0.5%, the depen- dence of an oblique angle to chamber axis have a large variance, take a example, when the beam orientation opposite to the tip of chamber along the axis of chamber, the reading reach to -5.44% comparing to the reading when beam perpendicular to chamber axis. The stem effect is less than 1% for Cobalt-60 gamma ray. For kilovolt X ray and Cobalt-60 gamma ray, the vari- ation of energy dependence response is less than 1%; The polarity effect is less than 0.2% for megavolt photon beam and elec- tron beam, but the recombination effect can't be ignored for electron beam, especially for 12 MeV and 15 MeV electron beam, they are 1.06% and 1.18%,respectively. Conclusion: The all technique index can meet the need of clinic work, so it can be used for clinic quality assurance, measurement and calibration.