Determination of Cisatracurium in Human Plasma by Iron-Pair HPLC with Fluorescence Detection

LI Juan,CHEN Bing,YANG Wan-hua,ZHANG Yu-long
DOI: https://doi.org/10.16155/j.0254-1793.2011.04.035
2011-01-01
Abstract:Objective:To establish iron-pair HPLC-fluorescence method for determination of cisatracurium in human plasma.Methods:Acidified plasma samples were pretreated with acetonitrile to precipitate protein and the supernatant liquid obtained from the plasma by centrifuge was subjected to chromatographic analysis.The separation was performed on a reversed-phase Agilent Eclipse C18(150 mm×4.6 mm,5 μm),the mobile phase was made up of 1.5% trifluoroacetic acid-acetonitrile(55∶45) at a flow rate of 0.5 mL·min-1,the excitation and emission wavelengths are 280 nm and 320 nm,respectively.Results:Calibration curves for cisatracurium was 12.4-4960.0 ng·mL-1(r=0.9998).The limit of detection was 1.24 ng·mL-1.The average extraction recovery was 85.1%-88.5%,and the average method recovery was 98.3%-102.1%.The intra-day and inter-day RSDs were both less than 3.4%.Conclusion:The proposed method is simple,sensitive,accurate and reliable for the need of clinical detection and the pharmacokinetic research.
What problem does this paper attempt to address?