High-Throughput Determination of Carbocysteine in Human Plasma by Liquid Chromatography/Tandem Mass Spectrometry: Application to A Bioequivalence Study of Two Formulations in Healthy Volunteers
HC Bi,LZ Zhao,GP Zhong,SF Zhou,B Li,Y Deng,X Chen,M Huang
DOI: https://doi.org/10.1002/rcm.2422
IF: 2.586
2006-01-01
Rapid Communications in Mass Spectrometry
Abstract:rapid and sensitive liquid chromatography/tandem mass spectrometry (LC/MS/MS) method to determine carbocysteine in human plasma was developed and fully validated. After methanol-induced protein precipitation of the plasma samples, carbocysteine was subjected to LC/MS/MS analysis using electrospray ionization (ESI). The MS system was operated in the selected ion monitoring (SRM) mode. Chromatographic separation was performed on a Hypurity C-18 column (i.d. 2.1 mm X 50 mm, particle size 5 mu m). The method had a chromatographic running time of 2.0 min and linear calibration curves over the concentration ranges of 0.1-20 mu g/ml, for carbocysteine. The lower limit of quantification (LLOQ) of the method was 0.1 mu g/ml, for carbocysteine. The intra- and inter-day precision was less than 7% for all quality control samples at concentrations of 0.5, 2.0, and 10.0 mu g/mL. These results indicate that the method was efficient with a simple preparation procedure and a very short running time (2.0 min) for carbocysteine compared with methods reported in the literature and had high selectivity, acceptable accuracy, precision and sensitivity. The validated LC/ MS/MS method has been successfully used to a bioequivalence study of two tablet formulations of carbocysteine in healthy volunteers. Copyright (c) 2006 John Wiley & Sons, Ltd.