Determination of Ondansetron in Rats Plasma by HPLC

Wenfeng Li,Wei Ma,Yang Yun-li,Fa-tuan Dong,PU Ying-jie
2011-01-01
Abstract:Objective To establish high-performance liquid chromatographic(HPLC)method for the quantification of ondansetron in rats plasma.Methods Sample preparation was accomplished through liquid-liquid extraction with ethyl acetate and chromatographic separation was performed on a reversed phase Waters Atlantis C18 column at 35 ℃.Mobile phase consisted of a mixture of acetonitrile-0.02 μL/mL KH2PO4 buffer(pH 3.3)(30:70,v/v)at a flow rate of 1.0 mL/min.Wavelength was set at 310 nm.Helicid was selected as the internal standard(IS).Results The linear range of ondansetron in plasma was 2~1000 ng/mL.The intra-day and inter-day precisions were lower than 5.1 and 7.7%,respectively.The recovery rate of ondansetron was 90.4~94.1%.Conclusion HPLC is rapid,sensitive and accurate,which can be used for determination of andansetron in rat plasma and its pharmacology study.
What problem does this paper attempt to address?