Research on Automatic Analyzer for Determining Trace-level Sodium Based on Flow-injection Ion-selective Electrode Method

李永生,杨莉玲,星成霞
DOI: https://doi.org/10.3969/j.issn.1002-1841.2004.03.011
2004-01-01
Abstract:A quick determination method for sodium ion of boiler water and steam at power plants with flow-injection analysis(FIA)and ion selective electrode(ISE)has been expoitured.The basiclize can be accomplished automatically,and the measurement is under sealed circumvent.The electrodes can be rinsed automatically;there is good linearity in the range of 2~100 μg/L.Also,it analyzs very quickly,50 samples/h can be analyzed and the consumption of sample is only 0.3mL/sample.It has been used to determine the trace-level sodium of boiler water and steam at power plants successfully.Alkalization reagent dosage is about 1/50 of general industrial sodium Instrument.The new pattern on-line sodium instrument based on it can reduce alkali dissipation,improve economy index,protect environment.
What problem does this paper attempt to address?