Determination of Anionic Surfactant by Intelligent Flow Injection and Ion Selective Electrode Analysis

魏福祥,王蕊,王浩然,王占辉
DOI: https://doi.org/10.3969/j.issn.0367-6358.2010.08.006
2010-01-01
Abstract:Based on virtual instrument technology,the flow injection analysis system used for determination of anionic surfactant in water was established.The results show that when double distilled water is used as carrier stream,the pump speed was 30 r/min the sampling volume was 260 μL and the length of disperse coil was 20 cm,the potential signal is stable.The linear range for the electrode is 1×10-6 ~1×10-3 mol/L,in which the electrode has a good Nernst response,and the detection limit is 4×10-7 mol/L.After amplification of the potential signal for 5 times,the response slope is 0.2806 V/dec,and the electrode has good stability and interference immunity.The spiked average recovery is 96.8%.