Design of Prototype Magnets for HUST Proton Therapy Beamline

Wei Chen,Bin Qin,Xiaoyu Fang,Qushan Chen,Jun Yang,Kaifeng Liu
DOI: https://doi.org/10.1109/tasc.2017.2775579
IF: 1.9489
2018-01-01
IEEE Transactions on Applied Superconductivity
Abstract:A proton therapy facility with two 360° gantry treatment rooms and one fixed beamline is under development at the Huazhong University of Science and Technology (HUST), which is based on an isochronous superconducting cyclotron. The energy is modulated in the range of 70-240 MeV with an energy selection system. In this paper, we report the main design specifications of the beamline in the HUST-PTF project, in which the main features are the intensity suppression scheme beamline. Two prototype magnets (one quadrupole and one dipole) used in the beamline are also studied and presented. Two-dimensional contour optimization and pole-end chamfer iteration are used to minimize the harmonic errors of the integral field. Finally, both the field homogeneity and multipole errors achieve the precision requirement over the operating field range.
What problem does this paper attempt to address?