PKU-PET-II: A novel SiPM-based PET imaging system for small animals

Zhaoheng Xie,Suying Li,Kun Zhou,Ivan Vuletic,Xiangxi Meng,Sihao Zhu,Huan Xu,Kun Yang,Baixuan Xu,Jinming Zhang,Qiushi Ren
DOI: https://doi.org/10.1016/j.nima.2017.09.012
2018-01-01
Abstract:The objective of this study was to introduce, describe, and validate the performance of a novel preclinical silicon photomultiplier (SiPM)-based PET system (PKU-PET-II). Briefly, the detector assembly consisted of cerium-doped lutetium–yttrium oxyorthosilicate (LYSO) crystals, with dimensions of 2×2×15 mm3, that offered a 60 mm transaxial field of view (FOV) and 32 mm axial FOV, respectively. The compact front-end electronics readout and digital controller implemented architecture in the FPGA were noteworthy improvements in PKU-PET-II over its predecessor (PKU-PET-I). Based on the National Electrical Manufacturers Association (NEMA) NU 04-2008 standards, the design of the PKU-PET-II system was validated by a phantom experiment. The results presented spatial resolution (evaluated as full width at half maximum) with a system range from 1.68±0.07 to 2.31±0.03 mm at the FOV center and from 1.43±0.02 to 2.10±0.10 mm at the 1/4th axial FOV, respectively. The system’s absolute sensitivity at the center position was 1.35% with the coincidence window of 6 ns and energy window of 300–700 keV. In addition, the NEMA image quality phantom and an animal study results validated the system imaging performance in preclinical imaging application. In conclusion, this SiPM-based, small-animal PET system (PKU-PET-II) provided higher-resolution, adequate sensitivity, and excellent image quality and has potential as a useful tool for real-time imaging of disease progression and development in vivo.
What problem does this paper attempt to address?