Abstract:Based on the direct absorption measurement of absorption spectrum of tunable diode laser absorption spectroscopy (TDLAS) technology, the DFB laser of 1 580 nm center wavelength is chosen to detect the concentration of CO2 in simulated flue gas at room temperature and atmospheric pressure.The concentration of CO2 is computed by removing absorption peak fitting method and pure N2 line fitting method, and the two computed results are compared.The results show that the maximum relative error of concentration is 2.64% for the pure N2 line fitting method, and the mean square value is 1.69%;the maximum relative error of concentration got from the removing absorption peak fitting method is 9.81%, and the mean-square value is 7.81%.Using the method of pure N2 line fitting method, the accuracy of concentration is greatly improved, so the method can provide baseline selection reference for CO2 measurement.